LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES D684-00-0479-01 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) D684-00-0479-02 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0479-01C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0479-02C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0479-07 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0479-08 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) D684-00-0479-09 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0479-10 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0479-01U LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0479-02U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0480-01 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0480-02 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0480-01C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0480-02C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0480-07 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) D684-00-0480-08 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0480-09 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0480-10 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0480-01U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0480-02U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES D684-00-0478-01 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) D684-00-0478-02 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (CHINA) D684-00-0478-01C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (CHINA) D684-00-0478-02C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (INDIA) D684-00-0478-07 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0478-08 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (TURKEY) D684-00-0478-09 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0478-10 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES, US ONLY D684-00-0478-01U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) US ONLY D684-00-0478-02U PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 25 cc IABs D884-00-0019-12 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 34/40 cc IABs D884-00-0019-13
Enforcement
- Recall Number
- Z-2223-2023
- Event ID
- 92527
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Datascope Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- August 2, 2023
- Initiation Date
- June 23, 2023
- Classification Date
- July 21, 2023
- Address
- 15 Law Dr Unit 1, N/A, Fairfield, NJ, 07004-3206, United States
Description
LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES D684-00-0479-01 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) D684-00-0479-02 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0479-01C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0479-02C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0479-07 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0479-08 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) D684-00-0479-09 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0479-10 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0479-01U LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0479-02U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0480-01 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0480-02 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0480-01C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0480-02C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0480-07 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) D684-00-0480-08 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0480-09 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0480-10 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0480-01U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0480-02U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES D684-00-0478-01 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) D684-00-0478-02 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (CHINA) D684-00-0478-01C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (CHINA) D684-00-0478-02C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (INDIA) D684-00-0478-07 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0478-08 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (TURKEY) D684-00-0478-09 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0478-10 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES, US ONLY D684-00-0478-01U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) US ONLY D684-00-0478-02U PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 25 cc IABs D884-00-0019-12 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 34/40 cc IABs D884-00-0019-13
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Product Description UDI LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES 10605767106557 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) 10605767106540 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10605767109329 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) 10605767109336 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10605767109695 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) 10605767109701 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) 10605767113180 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10605767113142 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10605767109503 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY 10605767109510 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES 10607567106564 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) 10607567106571 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109343 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) 10607567109350 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109718 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) 10607567109725 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113197 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10607567113159 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10607567109527 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, (APA) US ONLY 10607567109534 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES 10607567106526 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) 10607567106533 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (CHINA) 10607567109305 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (CHINA) 10607567109312 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (INDIA) 10607567109671 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (INDIA) 10607567109688 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (TURKEY) 10607567113203 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10607567113210 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES, US ONLY 10607567109480 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) US ONLY 10607567109497 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 25 cc IABs 10607567106632 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 34/40 cc IABs 10607567106649
Domestic distribution nationwide. International distribution worldwide.
321,609 total kits