FDA Enforcement
Class II
Terminated
Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
Recall: Z-2223-2017
·
Reported June 14, 2017
Enforcement
- Recall Number
- Z-2223-2017
- Event ID
- 77243
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- June 14, 2017
- Initiation Date
- May 9, 2017
- Classification Date
- June 2, 2017
- Termination Date
- February 2, 2018
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
Reason
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
Code Info
Serial Number 204C5690 204H0157 204H0158
Distribution
US distribution to WA only.
Quantity
3