FDA Enforcement Class II Terminated

Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Recall: Z-2220-2012 · Reported August 29, 2012

Enforcement

Recall Number
Z-2220-2012
Event ID
62649
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
DePuy Orthopaedics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 29, 2012
Initiation Date
July 24, 2012
Classification Date
August 17, 2012
Termination Date
April 30, 2013
Address
700 Orthopaedic Dr, N/A, Warsaw, IN, 46582-3994, United States

Description

Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Reason

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.

Code Info

Catalog number: 1221-36-054 and Lot Number: 231859

Distribution

Worldwide Distribution - USA including PA and the country of Canada

Quantity

20 liners