FDA Enforcement
Class II
Terminated
IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.
Recall: Z-2219-2019
·
Reported August 14, 2019
Enforcement
- Recall Number
- Z-2219-2019
- Event ID
- 83251
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ideal Implant Incorporated
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 14, 2019
- Initiation Date
- March 18, 2019
- Classification Date
- August 8, 2019
- Termination Date
- March 3, 2021
- Address
- 14881 Quorum Dr Ste 925, N/A, Dallas, TX, 75254-7132, United States
Description
IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.
Reason
The implants were deflating due to a piece of silicone disrupting the seal of the posterior valve shortly after implantation.
Code Info
Serial numbers between 0100292 and 0102777.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Country of Canada.
Quantity
1,704 implants