FDA Enforcement Class II Terminated

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Recall: Z-2219-2017 · Reported June 7, 2017

Enforcement

Recall Number
Z-2219-2017
Event ID
77260
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Exact Sciences Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 7, 2017
Initiation Date
April 28, 2017
Classification Date
June 1, 2017
Termination Date
August 11, 2017
Address
441 Charmany Dr, Madison, WI, 53719-1234, United States

Description

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Reason

Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

Code Info

Equipment ID / Serial Number: MIX-0051 / MX00002051, MIX-0052 / MX00002032, MIX-0053 / MX00002028, MIX-0054 / MX00002052, MIX-0055 / MX00002043, MIX-0056 / MX00002040, MIX-0057 / MX00002050, MIX-0058 / MX00002036, MIX-0059 / MX00002027, MIX-0060 / MX00002015, MIX-0061 / MX00002016, MIX-0063 / MX00002039, MIX-0064 / MX00002033, MIX-0065 / MX00002053, MIX-0066 / MX00002014,

Distribution

US to WI only

Quantity

15