FDA Enforcement Class II Terminated

cobas¿ Liat¿ Analyzer

Recall: Z-2216-2016 · Reported July 27, 2016

Enforcement

Recall Number
Z-2216-2016
Event ID
74094
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Molecular Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 27, 2016
Initiation Date
May 9, 2016
Classification Date
July 15, 2016
Termination Date
March 24, 2021
Address
1080 US Highway 202 S, N/A, Branchburg, NJ, 08876-3733, United States

Description

cobas¿ Liat¿ Analyzer

Reason

Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.

Code Info

US-IVD: M1-E-00383 & M1-E-00434

Distribution

Distributed in NC and WA.

Quantity

2