FDA Enforcement
Class II
Terminated
cobas¿ Liat¿ Analyzer
Recall: Z-2216-2016
·
Reported July 27, 2016
Enforcement
- Recall Number
- Z-2216-2016
- Event ID
- 74094
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Molecular Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- July 27, 2016
- Initiation Date
- May 9, 2016
- Classification Date
- July 15, 2016
- Termination Date
- March 24, 2021
- Address
- 1080 US Highway 202 S, N/A, Branchburg, NJ, 08876-3733, United States
Description
cobas¿ Liat¿ Analyzer
Reason
Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.
Code Info
US-IVD: M1-E-00383 & M1-E-00434
Distribution
Distributed in NC and WA.
Quantity
2