FDA Enforcement
Class III
Terminated
Sunquest Laboratory SMART & Sunquest Laboratory SMART Select
Recall: Z-2216-2012
·
Reported August 22, 2012
Enforcement
- Recall Number
- Z-2216-2012
- Event ID
- 62344
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Sunquest Information Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 22, 2012
- Initiation Date
- April 28, 2011
- Classification Date
- August 16, 2012
- Termination Date
- September 4, 2012
- Address
- 250 S Williams Blvd, Tucson, AZ, 85711-4472, United States
Description
Sunquest Laboratory SMART & Sunquest Laboratory SMART Select
Reason
In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (<Y>/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or
Code Info
Version 6.3.0 and Later
Distribution
Nationwide Distribution
Quantity
103 sites