00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."
Enforcement
- Recall Number
- Z-2215-2012
- Event ID
- 62540
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 22, 2012
- Initiation Date
- July 18, 2012
- Classification Date
- August 16, 2012
- Termination Date
- April 17, 2014
- Address
- 345 E Main St, Warsaw, IN, 46580-2746, United States
Description
00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
Part 00-4309-029-01: Lot 60535162, 60545338, 60576488, 60605855, 60637022, 60689253, 60772974, 60875277, 60918060, 60989817, 61011689, 61024161, and 61051024.
Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.
212 (179 USA)