Responder 5 is a nurse call communication system that connects patients with staff in a hospital. The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more. Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers
Enforcement
- Recall Number
- Z-2213-2013
- Event ID
- 65657
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Rauland-Borg Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 25, 2013
- Initiation Date
- May 10, 2013
- Classification Date
- September 17, 2013
- Termination Date
- May 6, 2015
- Address
- 1802 W Central Rd, N/A, Mount Prospect, IL, 60056-2230, United States
Description
Responder 5 is a nurse call communication system that connects patients with staff in a hospital. The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more. Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers
A counterfeited electronic component was identified in the products affected. The component, a data transceiver (transmitter/receiver), is associated with the intercommunication of these devices in the Responder 5 system across the L-Net data path. Failure of a data transceiver results in the loss of nurse call functionality in the associated room, and may disrupt communications on the L-Net run
1) Product Code 351000 (Branch Regional Controller); 2) Product Code 352000 (Corridor Light); 3) Product Code 352020 (Domeless Controller)
Worldwide distribution - US: (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NE, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and WV., and the countries of : Australia, Canada, Columbia, Lebanon, Mexico and Saudi Arabia.
Product Code 351000: 385 units; Product Codes 352000 and 352020: 14,808 units (total)