UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology Art.no. / REF no.: 15000004 Art.no. / REF no.: 15000006
Enforcement
- Recall Number
- Z-2212-2023
- Event ID
- 92534
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- UFSK - International OSYS Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 26, 2023
- Initiation Date
- May 31, 2023
- Classification Date
- July 20, 2023
- Address
- Kirchhoffstr. 1, N/A, Ratisbon, N/A, N/A, Germany
Description
UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology Art.no. / REF no.: 15000004 Art.no. / REF no.: 15000006
Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position
Art.no. / REF no.: 15000004 Serial number range: 2000- 2999 Art.no. / REF no.: 15000006 Serial number range: 3000-4506
Nationwide
239 chairs US