COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.
Recall: Z-2209-2012
·
Reported August 22, 2012
Enforcement
- Recall Number
- Z-2209-2012
- Event ID
- 62586
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Molecular Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 22, 2012
- Initiation Date
- February 1, 2012
- Classification Date
- August 15, 2012
- Termination Date
- July 16, 2013
- Address
- 1080 Us Highway 202 S, Branchburg, NJ, 08876-3733, United States
Description
COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.
Reason
Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce
Code Info
Lot Numbers: 11049503, 11053503, 11054503, 11525503 and 11526503.
Distribution
Nationwide Distribution (USA).
Quantity
2,132 boxes