FDA Enforcement
Class II
Ongoing
GEM FLOW COUPLER Monitor, PN 5156-00000-011
Recall: Z-2208-2023
·
Reported July 26, 2023
Enforcement
- Recall Number
- Z-2208-2023
- Event ID
- 92525
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Baxter Healthcare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 26, 2023
- Initiation Date
- June 19, 2023
- Classification Date
- July 20, 2023
- Address
- 1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States
Description
GEM FLOW COUPLER Monitor, PN 5156-00000-011
Reason
An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.
Code Info
GTIN 00844735006148, all serial numbers
Distribution
US
Quantity
122 units