FDA Enforcement Class II Ongoing

GEM FLOW COUPLER Monitor, PN 5156-00000-011

Recall: Z-2208-2023 · Reported July 26, 2023

Enforcement

Recall Number
Z-2208-2023
Event ID
92525
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 26, 2023
Initiation Date
June 19, 2023
Classification Date
July 20, 2023
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

GEM FLOW COUPLER Monitor, PN 5156-00000-011

Reason

An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.

Code Info

GTIN 00844735006148, all serial numbers

Distribution

US

Quantity

122 units