Sarns TCM; The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient The TCM with options will also supply water for cardioplegia, freeze water for an ice supply, monitor temperatures in the patient and extracorporeal circuit, and allow gradient rewarming relative to a venous blood temperature. Product Usage: The Sarns TCM is indicated for controlling and monitoring patient temperature.
Enforcement
- Recall Number
- Z-2208-2016
- Event ID
- 74389
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 27, 2016
- Initiation Date
- June 8, 2016
- Classification Date
- July 18, 2016
- Termination Date
- August 4, 2017
- Address
- 6200 Jackson Rd, N/A, Ann Arbor, MI, 48103, United States
Description
Sarns TCM; The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient The TCM with options will also supply water for cardioplegia, freeze water for an ice supply, monitor temperatures in the patient and extracorporeal circuit, and allow gradient rewarming relative to a venous blood temperature. Product Usage: The Sarns TCM is indicated for controlling and monitoring patient temperature.
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
The Sarns Temperature Control and Monitor unit (TCM), PIN 15747, Catalog Number 15747
Domestic: AK, AL, AR, AZ, CA , CO, CT, DC , DE , FL, GA, HI, IA, ID, IL, IN, KS, KY,LA , MA, MD, ME, MI , MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV , NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign: Argentina, Australia, Belgium,Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, England, Finland, France, Germany Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Malaysia Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Portugal, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates , Uruguay , Venezuela, Vietnam
6412 units