FDA Enforcement Class II Terminated

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.

Recall: Z-2208-2012 · Reported August 22, 2012

Enforcement

Recall Number
Z-2208-2012
Event ID
62568
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Healthcare Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 22, 2012
Initiation Date
July 11, 2012
Classification Date
August 15, 2012
Termination Date
January 23, 2017
Address
3000 Minuteman Road, Andover, MA, 01810, United States

Description

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.

Reason

Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

Code Info

Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114

Distribution

Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.

Quantity

12 units