FDA Enforcement
Class II
Terminated
Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.
Recall: Z-2208-2012
·
Reported August 22, 2012
Enforcement
- Recall Number
- Z-2208-2012
- Event ID
- 62568
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Healthcare Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 22, 2012
- Initiation Date
- July 11, 2012
- Classification Date
- August 15, 2012
- Termination Date
- January 23, 2017
- Address
- 3000 Minuteman Road, Andover, MA, 01810, United States
Description
Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.
Reason
Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.
Code Info
Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114
Distribution
Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.
Quantity
12 units