FDA Enforcement Class II Ongoing

Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DUREPAIR 5X8IN NCE, 62158.

Recall: Z-2207-2023 · Reported July 26, 2023

Enforcement

Recall Number
Z-2207-2023
Event ID
92514
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medtronic Neurosurgery
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 26, 2023
Initiation Date
June 13, 2023
Classification Date
July 20, 2023
Address
5290 California Ave, N/A, Irvine, CA, 92617-3073, United States

Description

Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DUREPAIR 5X8IN NCE, 62158.

Reason

Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory process, comparable to infection.

Code Info

REF/Expiration Date On or Prior/UDI-DI/Lots: 61100/March 31, 2025/00643169063808/2006039, 2007020, 2009028, 2010001, 2010016, 2102001, 2102020, 2103001, 2105028, 2106001, 2108011, 2108017, 2109022, 2110009, 2110030, 2111023, 2111045, 2112026, 2203020, 2204016, 2204035, 2207014, 2207046; 61105/February 28, 2025/00643169063815/2009060, 2011017, 2102011, 2102027, 2102047, 2103002, 2104042, 2105029, 2108001, 2108012, 2108016, 2109023, 2110010, 2111021, 2111024, 2111046, 2112027, 2202034, 2203034, 2204027, 2206008, 2207015, 2207033; 61106/December 31, 2024/00643169063822/2010017, 2103003, 2105030, 2105038, 2108013, 2108018, 2109004, 2110011, 2110031, 2111025, 2202035, 220302; 61110/March 31, 2025/00643169063839/2006039, 2010018, 2101014, 2102002, 2102028, 2103004, 2105031, 2108019, 2109024, 2110012, 2111026, 2201005, 2202036, 2204025, 2204040; 61111/April 30, 2025/00643169063846/2008039, 2010019, 2102021, 2103005, 2105032, 2106004, 2108014, 2110013, 2111027, 2202037, 2203022, 2203028, 2207016; 62100/January 31, 2025/00643169063853/2007021, 2008040, 2009026, 2009061, 2010014, 2011018, 2012001, 2101015, 2101020, 2102012, 2102022, 2102048, 2103019, 2104043, 2105033, 2106005, 2108020, 2109001, 2109020, 2109028, 2110014, 2110032, 2110033, 2111028, 2111035, 2111037, 2112001, 2112002, 2202038, 2203023, 2204001, 2204028, 2206013, 2206027, 2207017, 2207048, 2208009; 62105/January 31, 2025/00643169063860/2008005, 2008041, 2009029, 2010015, 2010020, 2012002, 2101016, 2102003, 2102013, 2102029, 2103006, 2103036, 2104044, 2105036, 2106035, 2108003, 2108015, 2109002, 2109003, 2110003, 2110015, 2110016, 2111022, 2111033, 2111047, 2201006, 2202014, 2202025, 2203015, 2203025, 2203031, 2204023, 2204036, 2204041, 2206014, 2206028, 2207018, 2207044; 62106/December 31, 2024/00643169063877/2008042, 2010021, 2101017, 2102023, 2103007, 2104045, 2105039, 2106007, 2108004, 2108021, 2109005, 2110017, 2110034, 2111029, 2202039, 2207019, 2207051; 62110/March 31, 2025/00643169063884/2006041, 2007022, 2008006, 2008043, 2010022, 2101009, 2101018, 2102026, 2102049, 2103020, 2104046, 2105001, 2105040, 2108005, 2108022, 2110004, 2110018, 2111036, 2201007, 2202015, 2202031, 2204017, 2204042, 2205031, 2205032, 2206007, 2207020, 2207049; 62111/May 31, 2025/00643169063891/2010023, 2012003, 2101019, 2102024, 2103008, 2105034, 2106009, 2108023, 2109006, 2110019, 2110035, 2111030, 2111034, 2202040, 2203024, 2204002, 2207021, 2207037, 2208027; 62158/April 30, 2025/00643169063907/2011019, 2103009, 2105035, 2110020, 2111031, 2202041, 2203033, 2204029, 2207006

Distribution

Worldwide - US Nationwide distribution including in the states of PR, TX, WA, HI, MI, GA, AL, IL, IN, IA, MO, KY, NY, CA, AZ, PA, UT, ME, MD, MN, FL, AK, NE, VA, OR, WI, NC, LA, SD, TN, MS, DE, NJ, ID, SC, OH, MA, DC, NM, WV, NH, CT, NV, OK, AR, KS, MT, RI, CO, ND, WY, VT and the countries of Barbados, Cayman Islands, Honduras, Belgium, Switzerland, Italy, United Kingdom, Canary Islands, Hungary, Poland, Germany, Spain, Czech Republic, Thailand, South Africa, Greece, Finland, Portugal, Colombia, Norway, Iraq, Jordan, Estonia, Austria, Slovakia, Libya, Qatar, Bosnia And Herzegovina, Algeria, Dominican Republic, Canada, Australia, New Zealand, Sweden, El Salvador, Ireland, United Arab Emirates, Saudi Arabia, Kazakhstan, Turkey, Mauritius, Ghana, Russian Federation, Lebanon, Romania, Iran, Islamic Republic Of, Egypt, Namibia, Senegal, Ecuador, Panama, Peru, Guyana, Mexico, Chile, Costa Rica, Viet Nam, Kuwait, Ukraine, Israel, Turkmenistan.

Quantity

65029