FDA Enforcement Class II Terminated

NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Recall: Z-2207-2020 · Reported June 10, 2020

Enforcement

Recall Number
Z-2207-2020
Event ID
85331
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nobel Biocare Usa Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 10, 2020
Initiation Date
July 19, 2019
Classification Date
June 2, 2020
Termination Date
October 27, 2020
Address
22715 Savi Ranch Pkwy, N/A, Yorba Linda, CA, 92887-4609, United States

Description

NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Reason

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

Code Info

Lot number: 13080792

Distribution

US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None

Quantity

221 implants