FDA Enforcement Class II Terminated

6.0mm Round Fluted Bur, Super Long

Recall: Z-2205-2016 · Reported July 20, 2016

Enforcement

Recall Number
Z-2205-2016
Event ID
74562
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Instruments Div. of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 20, 2016
Initiation Date
June 16, 2016
Classification Date
July 14, 2016
Termination Date
June 6, 2017
Address
4100 E Milham Ave, Portage, MI, 49002-9704, United States

Description

6.0mm Round Fluted Bur, Super Long

Reason

Stryker Instruments initiated a voluntary recall of specific lots of Round Fluted Burs, due to tarnishing or corrosion which may be present on the recalled burs and could result in a foreign body reaction (inflammation) necessitating surgical intervention.

Code Info

5190-010-060  Lot Numbers 15231017, 16033017, 16066017

Distribution

US Distribution to the states of : CA, IL, SD Foreign: Canada, Netherlands

Quantity

25