FDA Enforcement Class II Terminated

Carestream DRX-1 System Battery --- Made in U.S.A. for Carestream Health, Inc., Carestream Health, Inc., 150 Verona St, Rochester, NY 14608. --- The battery may be sold with the DRX-1 imaging device or as a stand-alone accessory.

Recall: Z-2202-2012 · Reported August 22, 2012

Enforcement

Recall Number
Z-2202-2012
Event ID
62310
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Carestream Health, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 22, 2012
Initiation Date
June 12, 2012
Classification Date
August 13, 2012
Termination Date
May 25, 2016
Address
1049 Ridge Rd W, Rochester, NY, 14615-2731, United States

Description

Carestream DRX-1 System Battery --- Made in U.S.A. for Carestream Health, Inc., Carestream Health, Inc., 150 Verona St, Rochester, NY 14608. --- The battery may be sold with the DRX-1 imaging device or as a stand-alone accessory.

Reason

Carestream DRX-1 System batteries manufactured before June 2012 due to two occurrences of lithium-ion battery overheating, which may cause burns.

Code Info

All DRX-1 batteries manufactured before June 2012.

Distribution

Worldwide Distribution, including Nationwide (USA).

Quantity

9971 units