FDA Enforcement Class II Ongoing

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Recall: Z-2198-2025 · Reported August 6, 2025

Enforcement

Recall Number
Z-2198-2025
Event ID
97184
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Capnia Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 6, 2025
Initiation Date
April 21, 2025
Classification Date
July 29, 2025
Address
1101 Chess Dr, Foster City, CA, 94404-1102, United States

Description

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Reason

Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.

Code Info

Monitor UDI-DI: 00858481006017. Lot (C20206): 231220

Distribution

US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.

Quantity

2160