FDA Enforcement Class II Terminated

BD Microtainer Quikheel Lancet 1.0 mm incision Depth 2.5 mm Incision Length REF 368101 Rx Only Becton, Dickinson and Company, Franklin Lakes, NJ 07417 USA Made in USA. Used to perform heel sticks for the purpose of collecting skin puncture blood.

Recall: Z-2192-2013 · Reported September 18, 2013

Enforcement

Recall Number
Z-2192-2013
Event ID
66031
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 18, 2013
Initiation Date
July 10, 2013
Classification Date
September 11, 2013
Termination Date
March 19, 2014
Address
1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States

Description

BD Microtainer Quikheel Lancet 1.0 mm incision Depth 2.5 mm Incision Length REF 368101 Rx Only Becton, Dickinson and Company, Franklin Lakes, NJ 07417 USA Made in USA. Used to perform heel sticks for the purpose of collecting skin puncture blood.

Reason

The Lancet will not retract after activation thus creating a potential for exposure to a used sharp.

Code Info

REF # 368101 Lot #3116007 Class I Exempt Device listing Number E158602

Distribution

Distributed in the states of CT, KY, IL, NY, PA, TX, and VA and the country of Canada.

Quantity

US - 39,800 units ; Canada - 18,400 units