FDA Enforcement Class II Ongoing

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Recall: Z-2189-2024 · Reported July 3, 2024

Enforcement

Recall Number
Z-2189-2024
Event ID
94644
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Draeger Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 3, 2024
Initiation Date
May 13, 2024
Classification Date
June 25, 2024
Address
6 Tech Dr, Andover, MA, 01810-2434, United States

Description

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Reason

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Code Info

Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23

Distribution

US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.

Quantity

34,102 units