FDA Enforcement Class II Ongoing

REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866

Recall: Z-2187-2021 · Reported August 11, 2021

Enforcement

Recall Number
Z-2187-2021
Event ID
88123
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Laboratories
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 11, 2021
Initiation Date
June 10, 2021
Classification Date
August 2, 2021
Address
1915 Hurd Dr, Irving, TX, 75038, United States

Description

REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866

Reason

There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

Code Info

Lot Numbers / Expiration Date 84126UN20 / 09JUL2021 03676UN20 / 25AUG2021 10597UN20 / 29AUG2021 26275UN20 / 20OCT2021 77748UN20 / 08DEC2021 65958UN20 / 08DEC2021 81824UN21 / 16FEB2022

Distribution

Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.

Quantity

2,212 units