Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000 Control Unit and Energy Transfer Pads is a thermal regulating system, indicated for monitoring and controlling patient temperature.
Enforcement
- Recall Number
- Z-2181-2012
- Event ID
- 62596
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Kimberly-Clark Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 15, 2012
- Initiation Date
- April 27, 2012
- Classification Date
- August 9, 2012
- Termination Date
- September 5, 2012
- Address
- 1400 Holcomb Bridge Rd, Roswell, GA, 30076, United States
Description
Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000 Control Unit and Energy Transfer Pads is a thermal regulating system, indicated for monitoring and controlling patient temperature.
A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.
Catalogue #01000-01 and 01000-01R, Units with 4-Digit Serial Numbers: Lot #4297, 4324, 4332, 4519, 4559, 4572; Units with 10 Character Serial Numbers: SA08017027, SA08017028, SA08017029, SA08017030, SA08017031, SA08017032, SA08017033, SA08017034, SA08017035, SA08017036, SA08017037, SA08017038, SA08017039, SA08017040, SA08017041, SA08017042, SA08017043, SA08017044, SA08017045, SA08017046, SA08017047, SA08017048, SA08017049, SA08343100, SA11137004, SA11137005, SA11137006, SA11137007, SA11137008, SA08094124, SA08094125, SA08094126, SA08094127, SA08094128, SA08094129, SA08094130, SA08094131, SA08094132, SA08094133, SA08094134, SA08094135, SA08094136, SA08094137, SA08094138, SA08094139, SA08094140, SA08094141, SA08094142, SA08094143, SA08094144, SA08094145, SA08094146, SA08094147, SA08094148, SA08094149, SA08094150, SA08094151, SA08094152, SA08094153, SA09020004, SA09020005, SA09020007, SA09020008, SA09020010, SA09020013, SA11298011, SA11298012, SA11298013, SA11298014, SA11298015, SA11298016, SA11298017, SA11298018, SA11298019, SA08136001, SA08136002, SA08136003, SA08136004, SA08136005, SA08136006, SA08136007, SA08136008, SA08136009, SA08136010, SA09043001, SA09043002, SA09043003, SA09043004, SA09043007, SA09043008, SA09043009, SA09043010, SA09043010, SA09043011, SA09043012, SA09043013, SA09043014, SA09043015, SA09043016, SA09043017, SA09043018, SA09043019, SA09043020, SA08343061, SA08343082, SA08343083, SA08343084, SA08343085, SA08343086, SA08343087, SA08343088, SA08343089, SA08343090, SA08343091, SA08343092, SA08343093, SA08343094, SA08343095, SA08343096, SA10337006, SA10337007, SA10337008, SA10337009 and SA10337010.
Nationwide Distribution-including the states of AZ, CA, CO, DE, FL, GA, IA, ID, IN, IL, KS, KY, LA, MA, MD, ME, MO, MS, NE, NM, OH, OR, PA, RI, SC, TN, TX, VA, WI and WV.
129 units