Target Detachable Coils: TARGET HELICAL NANO 2MM X 4CM MODEL Number:M0035452040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Enforcement
- Recall Number
- Z-2180-2016
- Event ID
- 74425
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Neurovascular
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 20, 2016
- Initiation Date
- June 6, 2016
- Classification Date
- July 11, 2016
- Termination Date
- March 3, 2017
- Address
- 47900 Bayside Pkwy, Fremont, CA, 94538-6515, United States
Description
Target Detachable Coils: TARGET HELICAL NANO 2MM X 4CM MODEL Number:M0035452040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
MODEL Number:M0035452040; Lot Number/Expiration Date: 18760131 31-Dec-18 18760129 31-Dec-18 18832208 31-Jan-19 18832819 31-Jan-19 18831597 31-Jan-19 18760132 31-Dec-18 18760130 31-Dec-18 18852438 28-Feb-19 18905625 31-Mar-19 18939389 31-Mar-19 19032961 30-Apr-19 19032868 30-Apr-19 19037908 31-May-19 18900809 28-Feb-19 18906580 28-Feb-19 18987592 30-Apr-19 19036655 31-May-19 18971469 30-Apr-19 18986652 30-Apr-19 18875322 28-Feb-19 19047456 31-May-19 18972020 30-Apr-19 18899967 28-Feb-19.
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
45,302 (8,188 in the U.S.) - total, all model numbers