Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.
Enforcement
- Recall Number
- Z-2179-2014
- Event ID
- 68888
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtronic MiniMed Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 20, 2014
- Initiation Date
- July 21, 2014
- Classification Date
- August 9, 2014
- Termination Date
- October 8, 2014
- Address
- 18000 Devonshire St, N/A, Northridge, CA, 91325-1219, United States
Description
Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.
Medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian monitors.
CSS7100 and CSS7100K
Worldwide:Distribution - US Nationwide and the countries of Croatia, Denmark, Finland, Germany, Netherlands, Norway, Sweden, Switzerland.
86 units