FDA Enforcement Class II Terminated

System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Recall: Z-2177-2012 · Reported August 15, 2012

Enforcement

Recall Number
Z-2177-2012
Event ID
62444
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 15, 2012
Initiation Date
July 26, 2012
Classification Date
August 9, 2012
Termination Date
November 16, 2012
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586, United States

Description

System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Reason

Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re

Code Info

catalog number: 801763 and serial numbers: 0011-0335, 0337-0610,1001-1034, 1037-1049,1100,1103-1106,1109-1450, and 1452.

Distribution

Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Japan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.

Quantity

992 units