FDA Enforcement Class I Ongoing

Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.

Recall: Z-2176-2025 · Reported August 13, 2025

Enforcement

Recall Number
Z-2176-2025
Event ID
97167
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
B BRAUN MEDICAL INC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 13, 2025
Initiation Date
July 9, 2025
Classification Date
August 6, 2025
Address
861 Marcon Blvd, N/A, Allentown, PA, 18109-9577, United States

Description

Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.

Reason

Label is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension set is not indicated for removal of air and does not feature an air vent, which is common to air eliminating filters.

Code Info

Catalog Number: V6215. Primary UDI-DI 04046964189173, Unit of Use UDI-DI: 04046964189166. Lot Number (Expiration Date): 0061742452 (30JUN2025), 0061747379 (31AUG2025 0, 0061767411 (31JAN2026), 0061780914 (30APR2026), 0061803499 (30SEP2026), 0061806173 (30NOV2026), 0061822333 (30APR2027), 0061836578 (30APR2027), 0061849109 (31JUL2027), 0061850836 (31AUG2027), 0061899902 (30SEP2028), 0061936368 (31MAR2029), 0061936119 (31MAR2029), 0061936369 (31MAR2029), 0061940584 (30APR2029).

Distribution

US distribution to AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WY. International distribution to Canada.

Quantity

34,825 units US