FDA Enforcement Class II Ongoing

The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Recall: Z-2176-2024 · Reported July 3, 2024

Enforcement

Recall Number
Z-2176-2024
Event ID
94700
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Maquet Cardiovascular, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 3, 2024
Initiation Date
May 17, 2024
Classification Date
June 24, 2024
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094, United States

Description

The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Reason

Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

Code Info

Model No. VH-3500, VH-3000-W; UDI: 00607567701250, 00607567700345; Lot No. 3000314373, 3000314821, 3000316023, 3000316024, 3000316027, 3000317501, 3000317575, 3000317576, 3000318302, 3000318901, 3000320140, 3000320141, 3000320773, 3000321483, 3000321723, 3000323214, 3000324397, 3000324409, 3000325013, 3000325249, 3000325689, 3000325776, 3000328206, 3000328561, 3000328562, 3000329723, 3000330445, 3000331131, 3000331388, 3000332428, 3000332510, 3000332993, 3000333524, 3000333829, 3000333956, 3000333967, 3000334315, 3000335670, 3000336708, 3000337272, 3000339701, 3000340534, 3000341133, 3000341675, 3000342250, 3000342251, 3000343264, 3000343266, 3000344887, 3000351187, 3000352770, 3000354226, 3000354228, 3000355224, 3000355369, 3000355798, 3000356048, 3000357837, 3000360574, 3000360576, 3000360775, 3000361036, 3000361423, 3000361707, 3000362064, 3000362258, 3000362671, 3000362933, 3000363162, 3000363947, 3000364082, 3000364309, 3000364520, 3000365379, 3000365391, 3000366446, 3000366831, 3000367160, 3000367439, 3000367542, 3000367852, 3000368165, 3000368658, 3000369005, 3000369122, 3000369645, 3000370069, 3000370290, 3000372257, 3000375534, 3000375776, 3000376132, 3000376388, 3000376542, 3000376883, 3000376883, 3000377320, 3000377699, 3000378079, 3000378654, 3000379282, 3000379691, 3000380264, 3000380475, 3000380671, 3000381287, 3000381620, 3000382054, 3000383367, 3000383779, 3000384088, 3000388588.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.

Quantity

37,186 units