FDA Enforcement Class II Ongoing

Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R

Recall: Z-2175-2025 · Reported August 6, 2025

Enforcement

Recall Number
Z-2175-2025
Event ID
97197
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Remote Diagnostic Technologies Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 6, 2025
Initiation Date
May 30, 2025
Classification Date
July 25, 2025
Address
Farnborough Aerospace Centre, Aerospace Boulevard, Farnborough, N/A, United Kingdom

Description

Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R

Reason

Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

Code Info

REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28

Distribution

Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.

Quantity

55