Dimension Vista(R) Vancomycin Flex(R) Reagent Cartridge (VANC, Catalog # K4086). For the quantitative measurement of vancomycin in human serum and plasma on the Dimension Vista system.
Enforcement
- Recall Number
- Z-2175-2012
- Event ID
- 62609
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 15, 2012
- Initiation Date
- July 17, 2012
- Classification Date
- August 9, 2012
- Termination Date
- January 14, 2014
- Address
- 500 GBC Drive, Mailstop 514, PO BOX 6101, Newark, DE, 19714-6101, United States
Description
Dimension Vista(R) Vancomycin Flex(R) Reagent Cartridge (VANC, Catalog # K4086). For the quantitative measurement of vancomycin in human serum and plasma on the Dimension Vista system.
Siemens has received complaints of customers experiencing slopes >1.05 when calibrating Flex(R) reagent cartridge lot 12095AD. This results in a slope failure and prevents the VANC assay from being run on the Dimension Vista(R) System.
lot 12095AD, exp 2013-04-04
Worldwide Distribution -- US, including Washington, DC and the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WV, and the countries of Canada, New Zealand, Germany, and Puerto Rico.
1,894 cartons