FDA Enforcement Class II Ongoing

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Recall: Z-2173-2024 · Reported July 3, 2024

Enforcement

Recall Number
Z-2173-2024
Event ID
94678
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SIGHT DIAGNOSTICS LTD
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 3, 2024
Initiation Date
May 13, 2024
Classification Date
June 21, 2024
Address
23, Derekh Menakhem Begin, Tel Aviv-Yafo, N/A, Israel

Description

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Reason

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

Code Info

REF: TK1, UDI-DI: B787TK10, Lot Code: 10114

Distribution

Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.

Quantity

7,450 units (2,800 US, 4,650 OUS)