FDA Enforcement Class II Terminated

Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria. This enzyme is characteristic of the genus Neisseria and most species of Pseudomonas. It enables the differentiation of Gram (-) bacilli (1, 2). A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.

Recall: Z-2172-2018 · Reported June 20, 2018

Enforcement

Recall Number
Z-2172-2018
Event ID
80111
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BioMerieux SA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 20, 2018
Initiation Date
March 23, 2018
Classification Date
June 12, 2018
Termination Date
June 23, 2020
Address
Chemin De L'Orme, N/A, Marcy L'Etoile, N/A, N/A, France

Description

Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria. This enzyme is characteristic of the genus Neisseria and most species of Pseudomonas. It enables the differentiation of Gram (-) bacilli (1, 2). A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.

Reason

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Code Info

Oxidase Reagent (ref: 55635): Lot number 92366601

Distribution

International distribution.

Quantity

4 boxes of OXIDASE REAGENT with 50 - 0.75ML bottles