1X1 Cleaving Tiles; The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient.
Enforcement
- Recall Number
- Z-2171-2015
- Event ID
- 71534
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- American Medical Systems Innovation Center - Silicon Valley
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 29, 2015
- Initiation Date
- May 28, 2015
- Classification Date
- July 21, 2015
- Termination Date
- January 19, 2016
- Address
- 3070 Orchard Dr, N/A, San Jose, CA, 95134-2011, United States
Description
1X1 Cleaving Tiles; The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient.
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
Model 80-10007-001; All product manufactured from April 8, 2012 to April 8, 2015
Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
16,710 total all affected devices