FDA Enforcement Class II Ongoing

CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient

Recall: Z-2169-2024 · Reported July 3, 2024

Enforcement

Recall Number
Z-2169-2024
Event ID
94494
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE Healthcare Finland Oy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 3, 2024
Initiation Date
April 5, 2024
Classification Date
June 21, 2024
Address
Kuortaneenkatu 2, Helsinki, N/A, Finland

Description

CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient

Reason

GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.

Code Info

GTIN 00195278522313: a) 5700023, Serial Numbers: 2534526-3-1, 2553213-5-1, 2606854-2-1, 5220256-1-1, 5215673-1-1, 5215634-1-1, 5264236-1-1, 5264113-1-1, 5245953-1-1, 5264089-1-1, 5275734-1-1, 5301535-1-1, 5300969-1-1, 5341313-1-1, 5395275-1-1, 5232613-1-1, 5232595-1-1, 5287222-1-1, 5390098-3-1, 5316614-6-1, 5316614-16-1, 5316614-22-1, 2615004-27-1, 5368586-1-1, 5223040-1-1, 5223040-1-2, 5223040-1-3, 5223040-1-4, 5274472-12-1, 5274472-71-1, 5350377-1-1, 5215675-1-1, 5223119-1-1, 5231395-1-1, 5245547-1-1, 5307272-2-1, 5318051-3-1, 5318978-2-1, 5317879-3-1, 5331606-1-1, 5338491-2-1, 5356848-4-1, 5356848-8-1, 5356848-7-1, 5356848-10-1, 5365520-6-1, 5373002-3-1, 5356866-18-1, 5356866-10-1, 5356866-26-1, 5356866-6-1, 5356866-2-1, 5374179-2-1, 2495499-1-1, 2495362-1-1, 2514941-1-1, 2522522-1-1, 2525199-1-1, 2520809-1-1, 2520809-1-2, 2526133-1-1, 2503865-2-1, 2514269-1-1, 2532289-1-1, 2534398-6-1, 2543443-1-1, 2538968-1-1, 2543938-15-1, 2547020-1-1, 2556233-14-1, 2556233-13-1, 2607079-5-1, 2556233-18-1, 2494479-1-1, 2494484-1-1, 2494487-1-1, 2515199-1-1, 2494485-1-1, 2494483-1-1, 2512827-1-1, 2512885-1-1, 2513077-1-1, 2518278-4-1, 2531556-1-1, 2525641-2-1, 2554943-1-1, 2551434-1-1, 2582151-1-1, 2585668-6-1, 2611315-1-1, 2611294-1-1, 2564890-3-1, 2619057-1-1, 2617112-1-1, 2618012-4-1, 2494648-5-1, 2500341-1-1, 2500318-2-1, 2509366-1-1, 2509367-1-1, 2512604-1-1, 2512604-1-2, 2514892-1-1, 2513939-2-1, 2516853-1-1, 2516851-1-1, 2523820-1-1, 2523643-2-1, 2524982-5-1, 2524982-1-1, 2537780-1-1, 2537780-1-2, 2522678-7-1, 2556763-18-1, 2556763-2-1, 2556763-18-2, 2562037-1-1, 2555090-2-1, 2576242-2-1, 2576259-2-1, 2554602-1-1, 2560382-1-1, 2560381-1-1, 2586284-1-1, 2612355-1-1, 2610438-1-1, 2611840-3-1, 2611351-2-1, 2611541-1-1, 2612895-4-1, 2604783-2-1, 2618900-1-1, 2618900-1-2, 2622551-1-1, 2623858-3-1, 2623105-1-1, 2623105-1-2, 2624426-2-1, 2626536-1-1, 2494497-1-1, 2494500-1-1, 2494525-1-1, 2503234-1-1, 2494526-1-1, 2495361-1-1, 2496099-1-1, 2516274-1-1, 2514690-1-1, 2495543-1-1, 2495400-1-1, 2495559-1-1, 2495409-1-1, 2519427-1-1, 2519427-1-2, 2519427-1-3, 2503764-1-1, 2522006-1-1, 2513114-3-1, 2514141-1-1, 2521852-1-1, 2512685-1-1, 2521540-1-1, 2522959-1-1, 2522761-1-1, 2519916-11-1, 2529981-1-1, 2529405-12-1, 2529406-11-1, 2529982-1-1, 2529409-9-1, 2529409-6-1, 2529434-1-1, 2543934-7-1, 2543933-7-1, 2543932-1-1, 2539102-7-1, 2539094-7-1, 2539436-1-1, 2540128-1-1, 2539084-8-1, 2539437-1-1, 2539438-1-1, 2528671-18-1, 2562722-1-1, 2568082-1-1, 2554514-1-1, 2547282-20-1, 2547282-6-1, 2547282-10-1, 2547282-8-1, 2547282-4-1, 2547231-10-1, 2547446-15-1, 2523479-16-1, 2552470-1-1, 2580467-1-1, 2561965-1-1, 2582525-6-1, 2582525-32-1, 2585579-2-1, 2592273-1-1, 2599020-1-1, 2579706-5-1, 2574524-13-1, 2562789-9-1, 2562789-2-1, 2562789-9-2, 2562789-5-1, 2562789-5-2, 2562789-4-1, 2562789-2-3, 2562789-37-1, 2562789-18-1, 2562789-14-1, 2562789-20-1, 2579723-1-1, 2580465-2-1, 2567049-24-1, 2580465-12-1, 2554559-19-1, 2604759-1-1, 2606427-1-1, 2575662-15-1, 2582438-1-1, 2612551-1-1, 2574524-13-2, 2611203-1-1, 2612612-10-1, 2611747-1-1, 2606474-1-1, 2611016-1-1, 2612734-18-1, 2612734-2-1, 2613185-15-1, 2613083-1-1, 2615930-1-1, 2613111-1-1, 2613111-20-1, 2618898-2-1, 2617993-2-1, 2617993-1-1, 2615926-1-1, 2616599-1-1, 2618691-30-1, 2616077-1-1, 2618690-20-1, 2621975-1-1, 2618903-10-1, 2622941-1-1, 2619537-1-1, 2621988-1-1, 2622940-1-1, 2620603-1-1, 2619182-2-1, 2625896-1-1, 2624865-1-1, 2624047-1-1, 2626740-2-1, 2624450-1-1, 2624725-18-1, 2624725-13-1, 2624725-5-1, 2624725-1-1, 2624725-13-2, 2577660-1-1, 2517501-1-1, 2545870-1-1, 2615937-3-1, 2616596-2-1, 2619984-2-1, 2616737-1-1, 2618481-5-1, 2499911-1-1, 2505712-1-1, 2503687-1-1, 2554926-1-1, 2572230-2-1, 2588785-1-1, 2495298-1-1, 2495264-1-1, 2509179-2-1, 2495358-1-1, 2515181-2-1, 2515181-1-1, 2513165-1-1, 2495299-1-1, 2531550-1-1, 2544853-2-1, 2574647-2-1, 2598884-1-1, 2618137-1-1, 2628573-1-1, 2494372-1-1, 2518561-9-1, 2522026-1-1, 2522834-1-1, 2524986-1-1, 2524986-7-1, 2524898-1-1, 2522845-1-1, 2529643-1-1, 2543789-1-1, 2552218-1-1, 2552156-1-1, 2552158-1-1, 2551090-7-1, 2557045-1-1, 2556176-13-1, 2571940-4-1, 2569835-2-1, 2574703-4-1, 2584917-4-1, 2595848-2-1, 2604561-1-1, 2604556-1-1, 2605946-1-1, 2605946-1-2, 2605946-1-3, 2605946-1-4, 2609030-1-1, 2611696-1-1, 2613465-1-1, 2612021-1-1, 2617132-1-1, 2616584-1-1, 2616584-1-2, 2619512-1-1, 2619506-1-1, 2618910-4-1, 2620627-1-1, 2620625-1-1, 2620672-1-1, 2619873-2-1, 2620665-1-1, 2619880-1-1, 2620617-1-1, 2620610-1-1, 2620667-1-1, 2620184-1-1, 2620624-3-1, 2620185-1-1, 2624861-1-1, 2625758-1-1, 2625946-1-1, 5365928-16-1, 2556663-2-1, 2556714-2-1, 5227069-1-1, 5272630-1-1, 5323410-2-1, 5363045-26-1, 2495391-1-1, 2495391-1-2, 2495390-1-1, 2495390-1-2, 2495390-1-3, 2495390-1-4, 2495391-1-3, 2495391-1-4, 2495388-2-1, 2521107-1-1, 2529146-1-1, 2546098-1-1, 2547269-1-1, 2547722-1-1, 2539431-1-1, 2547016-1-1, 2616640-1-1, 2620989-1-1, 2621646-1-1, 2618162-1-1, 2514419-2-1, 2514419-2-2, 5206450-1-1, 5154318-1-1, 5154276-1-1, 5206443-1-1, 5222668-1-1, 5300958-1-1, 5313048-22-1, 5313048-62-1, 5346596-31-1, 5344117-1-1, 5341616-12-1, 5341619-10-1, 2497074-1-1, 2515206-1-1, 2522537-1-1, 2515205-1-1, 2531024-2-1, 2531024-1-1, 2531180-1-1, 2551366-1-1, 2556063-4-1, 2534343-13-1, 2534343-17-1, 2532125-45-1, 2574438-1-1, 2589300-1-1, 2589311-1-1, 2589312-1-1, 2589301-1-1, 2589308-1-1, 2589309-1-1, 2578226-1-1, 2600361-23-1, 2586763-1-1, 2573863-2-1, 2586760-1-1, 2595286-1-1, 2586318-1-1, 2589302-1-1, 2589520-1-1, 2589511-1-1, 2586757-1-1, 2586753-1-1, 2586729-1-1, 2586734-1-1, 2586762-1-1, 2586752-1-1, 2605396-1-1, 2605342-1-1, 2605929-1-1, 2605935-1-1, 2605989-1-1, 2605951-1-1, 2610744-1-1, 2610826-18-1, 2610826-10-1, 2607292-1-1, 2608880-2-1, 2609315-1-1, 2609592-5-1, 2609559-1-1, 2609691-3-1, 2609340-1-1, 2609585-1-1, 2609334-1-1, 2609586-1-1, 2609350-1-1, 2609588-1-1, 2623578-16-1, 2622019-1-1, 2622025-1-1, 2622573-7-1, 2622573-27-1, 2622955-8-1, 2628534-3-1, 2628533-4-1, 2625902-6-1, 2495410-1-1, 2515674-1-1, 2514790-1-1, 2514715-1-1, 2515509-1-1, 2517461-1-1, 2519722-11-1, 2517647-1-1, 2516443-6-1, 2549822-22-1, 2549680-1-1, 2554101-1-1, 2562362-2-1, 2562825-4-1, 2571502-8-1, 2570126-2-1, 2577337-1-1, 2583607-6-1, 2570194-15-1, 2600414-14-1, 2605924-1-1, 2601688-1-1, 2601655-1-1, 2601803-1-1, 2611874-2-1, 2601748-1-1, 2601940-1-1, 2592691-6-1, 2592820-8-1, 2592691-7-1, 2592820-6-1, 2601920-1-1, 2613980-6-1, 2609421-4-1, 2601846-1-1, 2614641-11-1, 2610892-1-1, 2608694-2-1, 2615575-12-1, 2614969-22-1, 2615764-5-1, 2599224-2-1, 2621236-1-1, 2616938-3-1, 2616938-3-2, 2618511-6-1, 2624560-1-1, 2623606-5-1, 2626368-1-1, 2626370-1-1, 2626712-1-1, 2622224-9-1, 2623632-2-1, 2627177-8-1, 2623632-2-2, 2623632-2-3, 2627168-1-1, 2627801-36-1, 2628801-1-1, 2627695-9-1, 2626721-1-1, 2627175-2-1, 2627174-1-1, 2628301-1-1, 2509816-1-1, 2514797-1-1, 2517806-1-1, 2523476-1-1, 2523477-6-1, 2519756-3-1, 2520861-2-1, 2519763-4-1, 2519774-5-1, 2520365-3-1, 2519758-3-1, 2519758-5-1, 2523326-1-1, 2514456-4-1, 2524713-3-1, 2519512-1-1, 2519512-1-2, 2519512-1-3, 2528982-1-1, 2523876-4-1, 2525226-3-1, 2523871-1-1, 2523841-2-1, 2534939-1-1, 2524913-3-1, 2531974-1-1, 2533581-1-1, 2539845-3-1, 2537914-1-1, 2539820-8-1, 2535298-1-1, 2521153-4-1, 2536270-6-1, 2530263-5-1, 2530263-5-2, 2530263-1-1, 2560347-8-1, 2554261-6-1, 2554486-6-1, 2553619-8-1, 2553619-9-1, 2528652-7-1, 2528652-6-1, 2528652-9-1, 2576457-2-1, 2549815-32-1, 2549815-7-1, 2624547-1-1; b) 5700023-02, Serial Numbers: 5376661-1-1, 5376661-1-2, 5352820-3-1, 5266636-23-1, 5232164-5-1, 5232164-5-2, 5232640-38-1, 5232640-38-2, 5263649-44-1, 5233325-1-1, 5233325-1-2, 5263648-6-1, 5263648-6-2, 5263643-8-1, 5263646-12-1, 5261241-2-1, 5263645-19-1, 5263644-24-1, 4920193-12-1, 5296964-13-1, 5392029-1-1, 4861348-9-1, 5231103-28-1, 5231100-20-1, 5231106-2-1, 5231106-4-1, 5231104-39-1, 5231101-16-1, 5231102-4-1, 5231102-19-1, 5296738-19-1, 5296738-5-1, 5296738-1-1, 5296738-1-2, 5296738-4-1, 5231104-1-1, 5296738-35-1, 5354648-1-1, 5320993-14-1, 5320993-11-1, 5320993-44-1, 5097227-1-1, 5297079-14-1, 5335889-23-1, 5344243-4-1, 5344243-2-1, 5347576-2-1, 5359061-1-1, 5344696-1-1, 5288824-12-1, 5290581-5-1, 5272370-30-1, 5316921-1-1, 5233767-3-1, 5233767-2-1, 5252009-2-1, 5233165-47-1, 5377507-9-1, 5244310-2-1, 5251994-11-1, 5255034-4-1, 5255034-4-2, 5226930-18-1, 5241745-8-1, 5226930-4-1, 5226935-12-1, 5262866-2-1, 5262866-3-1, 5363729-19-1, 5212009-18-1, 5233042-24-1, 5233042-40-1, 5355390-3-1, 5291738-20-1, 5329207-2-1, 5354793-4-1, 5365513-18-1, 5368489-4-1, 5368492-13-1, 5318209-4-1, 5318202-11-1, 5318202-5-1, 5316711-13-1, 5316740-4-1, 5318188-15-1, 5316734-15-1, 5349390-3-1, 5374225-1-1, 5368169-34-1, 5368473-17-1, 5285417-4-1, 5354298-11-1, 5326282-8-1, 5330973-7-1, 5330973-3-1, 5361980-2-1, 5362007-54-1, 5232017-4-1, 5232017-4-2, 5300447-4-1, 5300278-4-1, 5300288-20-1, 5300295-12-1, 5338568-9-1, 5349366-1-1, 5349366-19-1, 5317581-29-1, 5331910-2-1, 5269037-8-1, 5269031-12-1, 5269015-25-1, 5028206-13-1, 5028206-13-2, 5169978-44-1, 5169997-44-1, 5169986-44-1, 5169950-1-1, 5364842-14-1, 5364842-3-1, 5262625-3-1, 5262623-80-1, 5374700-1-1, 5307843-20-1, 5371929-2-1, 5363162-10-1, 5356362-4-1, 5289111-19-1, 5353247-2-1, 5312936-16-1, 5312936-3-1, 5382031-1-1, 5382031-17-1, 5226855-3-1, 5292685-1-1, 5298493-13-1, 5298493-11-1, 5332148-3-1, 5332161-1-1, 5258798-11-1, 5258875-26-1, 5392201-1-1, 5392201-1-2, 5253414-1-1, 5331912-11-1, 5377458-1-1, 5336622-44-1, 5249468-2-1, 5249468-18-1, 5249458-15-1, 5249458-16-1, 5249669-13-1, 5249669-12-1, 2563777-1-1, 5353311-4-1, 5326776-1-1, 5223229-24-1, 5268496-9-1, 5267840-11-1, 5267840-13-1, 5268563-4-1, 5268563-2-1, 5268639-39-1, 5267932-24-1, 5271180-24-1, 5271522-24-1, 5356417-15-1, 5267884-24-1, 5373465-1-1, 5198791-1-1, 5266773-23-1, 5194983-1-1, 5194895-1-1, 5195392-1-1, 5194918-1-1, 5194589-1-1, 5390240-1-1, 5321832-50-1, 5250118-2-1, 2561736-1-1, 2561737-1-1, 2561738-1-1, 2575685-1-1, 5249743-1-1, 5250027-1-1, 5267969-1-1, 5250030-12-1, 5249745-1-1, 5284445-1-1, 5332331-1-1, 5249517-21-1, 5249592-1-1, 5250034-25-1, 5250032-16-1, 5307523-1-1, 5250130-1-1, 5250035-1-1, 5332362-1-1, 5336708-6-1, 5336708-14-1, 5332350-10-1, 2583232-1-1, 2588852-3-1; c) 5939590-102, Serial Numbers: 5377458-33377-240125A

Distribution

Worldwide distribution.

Quantity

4649 units