i-Stat cTnl cartridges Abbott Point of Care Inc. The i-Stat cTnl test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in heparinized whole blood or plasma samples. Cardiac troponin I measurements can be used as an aid in the diagnosis and treatment of myocardial infarction and in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality. The cartridge is to be used with the i-Stat Analyzer, but not with the i-Stat Portable Clinical Analyzer or the Philips Medical System (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-Stat System, the cTnl test is to be used by trained health care professionals in accordance with a facility's policies and procedures.

Recall: Z-2169-2012 · Reported August 15, 2012

Enforcement

Recall Number
Z-2169-2012
Event ID
62533
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Point Of Care Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 15, 2012
Initiation Date
February 1, 2012
Classification Date
August 9, 2012
Termination Date
February 8, 2013
Address
400 College Rd E, Princeton, NJ, 08540-6607, United States

Description

i-Stat cTnl cartridges Abbott Point of Care Inc. The i-Stat cTnl test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in heparinized whole blood or plasma samples. Cardiac troponin I measurements can be used as an aid in the diagnosis and treatment of myocardial infarction and in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality. The cartridge is to be used with the i-Stat Analyzer, but not with the i-Stat Portable Clinical Analyzer or the Philips Medical System (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-Stat System, the cTnl test is to be used by trained health care professionals in accordance with a facility's policies and procedures.

Reason

The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.

Code Info

List number 06F15-03, 06F15-04, 03P90-23, 600-9009-25 All lots

Distribution

Worldwide Distribution - USA including AZ and CO and internationally to Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.

Quantity

7,057,675 cartridges US ; 772,400 OUS