FDA Enforcement
Class I
Ongoing
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Recall: Z-2167-2025
·
Reported August 6, 2025
Enforcement
- Recall Number
- Z-2167-2025
- Event ID
- 97222
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Hamilton Medical AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 6, 2025
- Initiation Date
- June 26, 2025
- Classification Date
- July 30, 2025
- Address
- Parc Industrial Vial 10, Domat/Ems, N/A, Switzerland
Description
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Reason
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.
Code Info
UDI-DI: 07630002802963, Lots: 199675, 199676
Distribution
US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.
Quantity
2,560