Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Recall: Z-2167-2025 · Reported August 6, 2025

Enforcement

Recall Number
Z-2167-2025
Event ID
97222
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Hamilton Medical AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 6, 2025
Initiation Date
June 26, 2025
Classification Date
July 30, 2025
Address
Parc Industrial Vial 10, Domat/Ems, N/A, Switzerland

Description

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Reason

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

Code Info

UDI-DI: 07630002802963, Lots: 199675, 199676

Distribution

US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

Quantity

2,560