SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Enforcement
- Recall Number
- Z-2166-2017
- Event ID
- 76770
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 7, 2017
- Initiation Date
- January 31, 2017
- Classification Date
- May 26, 2017
- Termination Date
- May 24, 2018
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005, United States
Description
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8¿C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
Lot Numbers: 1) 357, 2) 358, 3) 360, 4) 361 & 5) 363 Expiration Dates: 1) 2017-01-30, 2) 2017-03-31, 3) 2017-04-30, 4) 2017-05-31 & 5) 2017-06-30
Nationwide Distribution
9100 kits