The VITEK 2 Neisseria-Haemophilus identification card (NH) is intended for use with the VITEK 2 Systems for the automated identification of most clinically significant fastidious organisms. The VITEK 2 NH identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
Enforcement
- Recall Number
- Z-2165-2018
- Event ID
- 80111
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- BioMerieux SA
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 20, 2018
- Initiation Date
- March 23, 2018
- Classification Date
- June 12, 2018
- Termination Date
- June 23, 2020
- Address
- Chemin De L'Orme, N/A, Marcy L'Etoile, N/A, N/A, France
Description
The VITEK 2 Neisseria-Haemophilus identification card (NH) is intended for use with the VITEK 2 Systems for the automated identification of most clinically significant fastidious organisms. The VITEK 2 NH identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
VITEK 2 NH (ref: 21346): Lot Number 2450355203
International distribution.
2 boxes of NH TEST KIT with 20 Cards