FDA Enforcement Class II Terminated

Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.

Recall: Z-2164-2013 · Reported September 18, 2013

Enforcement

Recall Number
Z-2164-2013
Event ID
64146
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
B Braun Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 18, 2013
Initiation Date
October 31, 2012
Classification Date
September 6, 2013
Termination Date
February 3, 2016
Address
1601 Wallace Dr Ste 150, N/A, Carrollton, TX, 75006-6690, United States

Description

Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.

Reason

This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removing DoseGuard dose limit protection. This occurs when a sequence of HOLD - HOLD key strokes are performed on the pump key panel while the user is in the Dose Dat

Code Info

Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES with main processor software versions prior to 151580. Multiple serial numbers affected. Affected serial numbers were notified of this field correction.

Distribution

Nationwide Distribution.

Quantity

1,970 pumps added for this expansion