Panther Fusion Extraction Reagent-S, Ref PRD-04331, IVD, CE, Rx Only consisting of FCR-S Fusion Capture Reagent - S, Nucleic Acid in a buffered solution and FER-S Fusion Enhancer Reagent-S (Lithium Hydroxide Solution), UDI: 15420045509214
Enforcement
- Recall Number
- Z-2163-2020
- Event ID
- 85346
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Hologic, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 3, 2020
- Initiation Date
- October 10, 2019
- Classification Date
- May 28, 2020
- Address
- 10210 Genetic Center Dr, N/A, San Diego, CA, 92121-4362, United States
Description
Panther Fusion Extraction Reagent-S, Ref PRD-04331, IVD, CE, Rx Only consisting of FCR-S Fusion Capture Reagent - S, Nucleic Acid in a buffered solution and FER-S Fusion Enhancer Reagent-S (Lithium Hydroxide Solution), UDI: 15420045509214
Reagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to invalid runs. This could result in incorrect results being reported and a delay in treatment.
Panther Fusion Extraction Reagent-S Lot number: 258402 Panther Fusion Capture Reagent-S Lot number: 257472 Panther Fusion Enhancer Reagent-S Lot number: 257473
US: AR, CA, FL, ID, IL, IN, LA, NM, NY, PA, TX, UT, VI, WA, WI OUS: AU, CA
47 kits