FDA Enforcement Class II Terminated

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Recall: Z-2160-2019 · Reported August 14, 2019

Enforcement

Recall Number
Z-2160-2019
Event ID
83345
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 14, 2019
Initiation Date
June 18, 2019
Classification Date
August 5, 2019
Termination Date
May 22, 2020
Address
2 Pearl Ct, N/A, Allendale, NJ, 07401-1611, United States

Description

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Reason

Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.

Code Info

Lot # B88751 UDI: (01)07613327351576(11)190125(10)B88751

Distribution

US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX

Quantity

11