FDA Enforcement Class II Terminated

Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 Product Usage: Intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, and bone and soft tissue reconstruction in pediatric patients and adults.

Recall: Z-2160-2017 · Reported May 31, 2017

Enforcement

Recall Number
Z-2160-2017
Event ID
77148
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 31, 2017
Initiation Date
April 26, 2017
Classification Date
May 24, 2017
Termination Date
February 7, 2020
Address
325 Corporate Dr, Mahwah, NJ, 07430-2006, United States

Description

Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 Product Usage: Intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, and bone and soft tissue reconstruction in pediatric patients and adults.

Reason

Stryker is recalling Hoffmann LRF (Limb Reconstruction Frame) System Safety Clip - Bone Transport Strut because there is a potential for loosening of the safety clips from the intended position on the Quick Release mechanism of the Transport Struts.

Code Info

Lot Numbers: J49236, V07241, V30655, V47575 and V49404

Distribution

US Nationwide distribution in the states of CA, FL, MD and VT

Quantity

437 units