DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013 Product Usage: The Salivary Testosterone ELISA (SlV-3013) is an enzymatic in vitro Assay for the detection of Testosterone in human Saliva. Testosterone levels are measured in fertility studies in men and women. Fertility problems are not diagnosed on the outcome of Testosterone alone but accompanied by other diagnostic means
Enforcement
- Recall Number
- Z-2159-2019
- Event ID
- 83183
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DRG Instruments GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 14, 2019
- Initiation Date
- May 21, 2019
- Classification Date
- August 5, 2019
- Termination Date
- May 29, 2020
- Address
- Frauenbergstr. 18, N/A, Marburg, N/A, N/A, Germany
Description
DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013 Product Usage: The Salivary Testosterone ELISA (SlV-3013) is an enzymatic in vitro Assay for the detection of Testosterone in human Saliva. Testosterone levels are measured in fertility studies in men and women. Fertility problems are not diagnosed on the outcome of Testosterone alone but accompanied by other diagnostic means
Salivary assay may observe a higher percentage of samples with 0 pg/ml testosterone or results which are too low. Saliva samples in the low measuring range of the kit(< 50 pg/ml) were assayed significantly lower
Lot Numbers: 66K118, 66K019
Worldwide Distribution - US Nationwide CA, MI Foreign: Canada, Brazil, China, Poland
38 kits