FDA Enforcement Class III Terminated

PTS Panels Lipid Panel test strip: IVD Lipid Panel Test Strips Ref Number: 1710 Product Usage: PTS Panels Lipid Panel test strips measure total cholesterol, HDL (high density lipoprotein) cholesterol, and triglycerides in whole blood with the CardioChek PA or the CardioChek Plus professional analyzers, and provide a quantitative result. A TC/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA and CardioChek Plus analyzers. A MEMo Chip¿ is provided with each package of test strips and must be properly inserted into the analyzer before any test can be run. The MEMo Chip contains the test name, calibration curve, lot number and test strip expiration date. After the test strip is inserted into the analyzer and blood applied to the test strip, test results are displayed in as little as 90 seconds.

Recall: Z-2157-2019 · Reported August 14, 2019

Enforcement

Recall Number
Z-2157-2019
Event ID
83097
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Polymer Technology Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 14, 2019
Initiation Date
June 7, 2019
Classification Date
August 5, 2019
Termination Date
July 8, 2020
Address
7736 Zionsville Rd, N/A, Indianapolis, IN, 46268-2175, United States

Description

PTS Panels Lipid Panel test strip: IVD Lipid Panel Test Strips Ref Number: 1710 Product Usage: PTS Panels Lipid Panel test strips measure total cholesterol, HDL (high density lipoprotein) cholesterol, and triglycerides in whole blood with the CardioChek PA or the CardioChek Plus professional analyzers, and provide a quantitative result. A TC/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA and CardioChek Plus analyzers. A MEMo Chip¿ is provided with each package of test strips and must be properly inserted into the analyzer before any test can be run. The MEMo Chip contains the test name, calibration curve, lot number and test strip expiration date. After the test strip is inserted into the analyzer and blood applied to the test strip, test results are displayed in as little as 90 seconds.

Reason

The MEMo chips provided with lot P901 may be mislabeled, the MEMo Chip may display Q901 on the analyzer LCD screen, instead of P901

Code Info

Lot Number: P901 Exp. Date: 8/17/2020 (01) 00381931710016 (17) 200817 (10) P901

Distribution

Worldwide Distribution - US Nationwide Bahamas Belgium Bermuda Brazil England Indonesia Latvia Mexico Nigeria Puerto Rico South Africa Sweden Taiwan

Quantity

1,887 vials US; 2,728 vials OUS