FDA Enforcement Class II Terminated

Sureflex 200 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection and incision of soft and cartilaginous tissue and surgical procedures involving vaporization, ablation and fragmentation of calculi (urinary and biliary).

Recall: Z-2157-2015 · Reported July 29, 2015

Enforcement

Recall Number
Z-2157-2015
Event ID
71534
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
American Medical Systems Innovation Center - Silicon Valley
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 29, 2015
Initiation Date
May 28, 2015
Classification Date
July 21, 2015
Termination Date
January 19, 2016
Address
3070 Orchard Dr, N/A, San Jose, CA, 95134-2011, United States

Description

Sureflex 200 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection and incision of soft and cartilaginous tissue and surgical procedures involving vaporization, ablation and fragmentation of calculi (urinary and biliary).

Reason

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex'" Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

Code Info

Model R-LLF200TG All product manufactured from April 8, 2012 to April 8, 2015

Distribution

Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.

Quantity

~16, 710 all affected devices