FDA Enforcement Class II Ongoing

Proteus 235-Proton Therapy System

Recall: Z-2155-2021 · Reported August 4, 2021

Enforcement

Recall Number
Z-2155-2021
Event ID
88196
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 4, 2021
Initiation Date
May 26, 2021
Classification Date
July 28, 2021
Address
Chemin Du Cyclotron 3, N/A, Louvain-la-neuve, N/A, N/A, Belgium

Description

Proteus 235-Proton Therapy System

Reason

if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment field may be delivered with an error in position

Code Info

Serial Numbers: PAT.107 (EU), PAT.108 (US), PAT.109 (US), PAT.115 (EU), SAT.117 (EU), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.125 (IN), SAT.126 (US), SAT.132 (EU), SAT.133 (US).

Distribution

FL, PA, TN, TX, VA Foreign: Germany, Italy, Russia, Sweden , India, Netherlands

Quantity

5 US; 8 OUS