FDA Enforcement
Class II
Ongoing
Proteus 235-Proton Therapy System
Recall: Z-2155-2021
·
Reported August 4, 2021
Enforcement
- Recall Number
- Z-2155-2021
- Event ID
- 88196
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 4, 2021
- Initiation Date
- May 26, 2021
- Classification Date
- July 28, 2021
- Address
- Chemin Du Cyclotron 3, N/A, Louvain-la-neuve, N/A, N/A, Belgium
Description
Proteus 235-Proton Therapy System
Reason
if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment field may be delivered with an error in position
Code Info
Serial Numbers: PAT.107 (EU), PAT.108 (US), PAT.109 (US), PAT.115 (EU), SAT.117 (EU), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.125 (IN), SAT.126 (US), SAT.132 (EU), SAT.133 (US).
Distribution
FL, PA, TN, TX, VA Foreign: Germany, Italy, Russia, Sweden , India, Netherlands
Quantity
5 US; 8 OUS