enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD
Enforcement
- Recall Number
- Z-2143-2016
- Event ID
- 74354
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ortho-Clinical Diagnostics
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 20, 2016
- Initiation Date
- May 16, 2016
- Classification Date
- July 8, 2016
- Termination Date
- June 8, 2018
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101, United States
Description
enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD
If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text to the LIS. The IFU states the interpretation term for these numerical result values is Retest?.¿ The difference in terminology exists for US Markets only.
Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; Serial Numbers/J Numbers: JIM341866, JIM341820, JIM341679, JIM341352, JIM343421, JIM341174, JIM336745, JIM340667, JIM338283, JIM-341592, JIM336957, JIM341996, JIM342334, JIM343694, JIM336855, JIM334236, JIM341415, JIM341392, JIM336481, JIM340934, JIM336167, JIM342249, JIM338290, JIM341750, JIM341426, JIM334986, JIM334527, JIM340024, JIM341021, JIM340658
Nationwide
US: 30 units