FDA Enforcement Class I Terminated

AUTOMIX 3+3 Compounder with Accusource Monitoring System, product 2M8287; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.

Recall: Z-2142-2012 · Reported August 15, 2012

Enforcement

Recall Number
Z-2142-2012
Event ID
62558
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Baxter Healthcare Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 15, 2012
Initiation Date
July 13, 2012
Classification Date
August 8, 2012
Termination Date
August 11, 2017
Address
Rt. 120 & Wilson Rd, Round Lake, IL, 60073, United States

Description

AUTOMIX 3+3 Compounder with Accusource Monitoring System, product 2M8287; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.

Reason

After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.

Code Info

Product code 2M8287, all serial numbers

Distribution

Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.

Quantity

950 units