FDA Enforcement Class II Ongoing

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Recall: Z-2141-2024 · Reported June 19, 2024

Enforcement

Recall Number
Z-2141-2024
Event ID
94676
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Fresenius Medical Care Holdings, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 19, 2024
Initiation Date
May 16, 2024
Classification Date
June 13, 2024
Address
920 Winter St Bld 920, N/A, Waltham, MA, 02451-1521, United States

Description

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Reason

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

Code Info

UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087

Distribution

US Nationwide distribution in the states of AL, IN, AK, WI.

Quantity

4 units