lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.
Enforcement
- Recall Number
- Z-2140-2018
- Event ID
- 80096
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Alere San Diego Inc. DBA Immunalysis Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 20, 2018
- Initiation Date
- November 28, 2017
- Classification Date
- June 8, 2018
- Termination Date
- January 29, 2021
- Address
- 829 Towne Center Dr, N/A, Pomona, CA, 91767-5901, United States
Description
lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.
The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.
Part Number: 3006-5, Lot Number: EK16135
US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV
37 kits