FDA Enforcement Class III Terminated

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.

Recall: Z-2140-2018 · Reported June 20, 2018

Enforcement

Recall Number
Z-2140-2018
Event ID
80096
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Alere San Diego Inc. DBA Immunalysis Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 20, 2018
Initiation Date
November 28, 2017
Classification Date
June 8, 2018
Termination Date
January 29, 2021
Address
829 Towne Center Dr, N/A, Pomona, CA, 91767-5901, United States

Description

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.

Reason

The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.

Code Info

Part Number: 3006-5, Lot Number: EK16135

Distribution

US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV

Quantity

37 kits